Autolyzer 800
Autolyzer 800
Autolyzer 800 is a full auto biochemistry analyzer capable to perform detailed and automated tests on whole blood, serum, plasma, cerebrospinal fluid or urine samples to determine concentration of analytes of Routine Biochemistry, Electrolytes, Enzyme etc. Simple on-board step by step instructions ensures consistency of results on which you can rely upon.
Technical Specification
System Function
- Throughput: 800 tests/hour
- 1200 tests/h with ISE( K+, Na+, Cl-)
- Contents of Testing: Clinical chemistry. Immunoturbidimetric
- Analytical Modes: Endpoint, Two-point, Kinetic
- Programming: Open system with user defined programs
- Automatic, Random Access
- STAT sample priority
Sample Tray
- Sample Position: 140 sample positions
- Sample Volume: 3~45 μL. 0.5 μL adjustable
- Inner sample tray
Reagent Tray
- Reagent Position: 2 reagent trays, 128 reagent positions
- Onboard refrigerated storage for reagent
- Reagent Volume
R1: 180~450 μL. 1 μL adjustable
R2: 30~250 μL. 1 μL adjustable
Reaction Tray
- Reaction Positions: 90 cuvettes, 10 cuvettes/strip
- Cuvette: Optical length 5mm
- Reaction Volume: 180~500ul
- Reaction Temperature: 37 °C, fluctuation: ±0.4 °C
- Simultaneously run 60 items (2 reagents) on board
Probe
- Pre-heating for Reagent
- Collision protection, Liquid level detection and Inventory chocking
- Automatic inner and outer probe washing alter every cycle, carry-over <0.1%
- 8-step washing system
Print out
- Built-in thermal printer, external printer optional
Measuring and Optic System
- Lamp: Halogen lamp
- Absorbance Range: 0~3.500Abs±1%
- Wavelengths: 12 wavelengths from 340-800nm
- Resolution: 0.0001ABS
Working Condition
- Power Supply: 100-240 VA 50/60 Hz
- Temperature: 10~35 °C
- Humidity: 35~80%
- Water Consumption: 1.5 L/hour
- Dimension: 583cm(L)x453cm(W)x530cm(H)
- Net weight. 70 KG